CHN Analyzer
Product Overview
Specification:-
Field Application โ Coal / Coke and other carbonaceous rock material
Conforming to the international standard ASTM D 5373.
Carbon, Hydrogen, Nitrogen must be detected using Infrared / Thermal Conductivity Detector.
Instruments must have the facility forโฆ
Description
Specification:-
- Field Application โ Coal / Coke and other carbonaceous rock material
- Conforming to the international standard ASTM D 5373.
- Carbon, Hydrogen, Nitrogen must be detected using Infrared / Thermal Conductivity Detector.
- Instruments must have the facility for internal calibration by analyzing standard sample.
- Balance โ having accuracy of 0.1 mg. with automatic weight entry facility to
analyzer/equipment.
- Instrument should be capable of analyzing sample portion of 50 mg to 200 mg of 212 micron size coal/coke sample.
- Power to the instruments โ 230 ยฑ 10% V AC, 50Hz.
- System must have fully automatic operation facility at user command basis. Auto sampler should have 30 or more positions.
- Tender submitted must carry supportive documents for the above specifications and range of detection of above elements repeatability/accuracy as per international standard etc., with branded computer ( Dell, HP, Compaq, Lenovo etc.) having following specifications- Processor i7 or better, 500 GBHDD, 4 GB RAM, 19โ LED colour monitor with USB ports, multimedia keyboard, optical mouse and LaserJet printer.
- Supplier should provide at least one certified reference coal sample with certified values of C, H, N.
- Gas, Cylinders & Regulators โ Equipment should have the capability to use Helium gas as
carrier gas and Oxygen as combustion gas. Two gas filled (He & O2) 47 liter water capacity steel cylinder and two double stage SS regulators (BIS certified) should be supplied with equipment (Price should be quoted separately).
- Tender submitted must carry the copy of recent 2 to 3, purchase order for similar instrument.
- Warranty โ 5 years comprehensive for complete instruments
- Consumable for testing of 5000 sample must be supplied with the instrument. It should be quoted separately.
- Others โ Installation, training and demonstration on samples for a minimum period of five days or till satisfaction whichever is greater.