| Mode |
General |
Serum |
Urine |
| Range |
Na: 0.1-100 ppm
Upto 250 meq/l, 1:100 dilK: 0.1-100 ppm
Upto 250 meq/l, 1:100 dilCa: 15-100 ppm
Upto 250 meq/l, 1:100 dilLi: 0.5-100 ppm
Upto 250 meq/l, 1:100 dil |
Na: 0-200
1:100 dilK: 0-10
1:100 dilCa: - Li: 0-2
1:5 dil |
Na: 0-250
1:100 dilK: 0-100
1:100 dilCa: - Li: 0-10
1:2 dil |
| Sensitivity |
General Mode: Na:0.5 ppm, K:0.5 ppm, Li:0.5 ppm, Ca:15 ppm |
| Filters |
Na & K Supplied (Ca & Li are optional at extra cost) |
| Resolution |
0.1 ppm/meq |
| Reproducibility |
+ 2% FS, &± 2 Digits |
| Curve Fit Accuracy |
+ 2% FS |
| Display |
20 x 4 Alphanumeric backlit LCD display |
| Average Time |
In-built in Software |
| Flame System |
LPG & dry oil free air |
| Detector |
Silicon Photodiode |
| Calibration |
Upto 5-Point Calibration with curve fitting software |
| Gas Control |
Adjustable with knobs |
| Ignition System |
Auto Ignition |
| Automizer |
Axial flow type |
| RS-232C Interface |
Available (PC Software at extra cost) |
| Power |
230 V + 10% AC, 50 Hz |
| Dimensions |
365 x 245 x 220 mm (L x B x H) (Approx.) |
| Weight |
Main Unit : 7.5 Kg. (Approx.) |